Using nuclear magnetic resonance to test fat content

A contract laboratory dramatically increased sample throughput using nuclear resonance technology.

The quickening pace of business caused one of Europe's leading independent testing laboratories to seek an alternative to the standard solvent extraction/acid hydrolysis (Soxhlet) method for testing the amount of fat in a variety of foods. This contract laboratory, with numerous sites throughout the UK and Ireland, provides quality control analytical services to the food industry. By converting from the wet chemistry method to an MQC benchtop nuclear magnetic resonance (NMR) analyzer for measuring fat content of foods, the lab reaped significant economic and environmental benefits. Fig.1. is a brief overview of NMR's advantages over other secondary methods.

Customers send foodstuff samples to this contract quality control laboratory, which has a specialty in fast turnaround service. A typical request includes five or six measurements, including fat (oil) content.

The Soxhlet method used for the oil measurement is slow, with measurements taking as long as six hours. This situation led to serious bottlenecks that were reducing throughput and affecting the lab's ability to deliver its promised rapid analysis service.

The process is also rather cumbersome, can be inaccurate, and requires highly skilled personnel. In addition, many of the hazardous chemicals used are becoming increasingly unacceptable according to international environmental standards.

The lab began seeking a rapid technique that would improve turnaround time without increasing operating costs, but would also be comparable to the industry standard Soxhlet technique.[Page Break]

Analytical methods

There are a number of analytical methods that can be used to conduct the testing. Such methods are often referred to as secondary techniques, since they are usually set-up to match the results produced by solvent extraction.

To provide a result equivalent to the traditional extraction techniques, secondary techniques must be correlated against the reference technique used.

Although they may be fast in use, many secondary techniques need to be calibrated and maintained regularly. Maintenance and consumables add significantly to the cost of ownership.

For example, Supercritical Fluid Extraction (SFE) is reasonably fast, but it requires high maintenance and the cost of compressed CO2 that is used to extract oil is also significant. Near Infra-Red (NIR) is sometimes used, but it is generally sensitive to the surface rather than the bulk of the sample, and has substantial calibration and calibration maintenance issues.[Page Break]

Product granularity

NIR calibration is complex because measurements are sensitive to product granularity and other physical characteristics and can be affected by additives such as seasoning, making it difficult to maintain accurate calibrations on a large variety of product types. This gives NIR limited applicability for the quality control of fat content in foodstuffs.

In contrast to the standard wet chemistry methods and various secondary techniques, low field nuclear magnetic resonance (NMR) provides a fast, direct and user friendly method for determination of the fat and oil content in foodstuffs.

The technique is based on measurement of the NMR response obtained from fat in the product, and quantification of the fat content by simple and direct calibration without the use of chemometrics. The instrument is extremely easy to operate and does not require the use of skilled chemists or NMR specialists.

NMR can be calibrated to cover a concentration range from 0.5 to 100 per cent fat. The user can produce a primary calibration using a single sample of fat. NMR is very stable over the long-term, and therefore requires infrequent recalibration.

Sample measurement time is short, typically about 20 seconds, allowing a high throughput of samples and efficient laboratory operation. Minimal sample preparation is required because the entire sample is normally loaded into a tube and measured directly, and there are several different size tubes available.

With NMR, no solvents are required since the sample is analysed in its natural state. The instrument facilitates bulk measurement; the signal is generated from the whole sample, ensuring that the result embodies everything inside the sample, not just on the surface. NMR is virtually insensitive to sample granularity and such additives as spices, flavors, colors and salt.

Finally, unlike Soxhlet, the NMR technique is non-destructive, so any required repeatability measurements can be made easily.[Page Break]

Throughput challenges

After reviewing the potential alternative solutions to bottlenecks associated with fat measurements, the laboratory contacted Oxford Instruments, which offers a benchtop NMR instrument widely used in industry to measure the oil content in foodstuffs and oil seeds. Oxford Instruments recommended its powerful but compact MQC benchtop NMR analyser for this application because it offers the analyst the benefits of accurate quantitative results, combined with sampling ease and convenience.

To verify whether the MQC NMR instrument would meet the laboratory's needs, applications specialists from Oxford Instruments collected and tested samples of some of the foods the lab typically analyses and compared NMR measurements with fat measurement values obtained with the Soxhlet method.

The goals of the testing process were to analyze 80 percent of samples using the MQC NMR instrument, achieve correlation to within 5 per cent of the wet chemistry method, and achieve a repeatability of within 5 percent.

Applications specialists sampled a range of foods, with fat contents ranging from 2.1 to 40.2 percent by mass, including baked cheese, muesli, milk powder, chicken powder, trifle, garlic bread, macaroni and cheese, meat, and chicken sandwich filler. Sampling was conducted using an Oxford Instruments MQC-23 benchtop NMR analyser equipped with a 26millimetre (mm) diameter probe.

The NMR results compared very closely to wet chemistry.

Fig. 2 shows a calibration for the samples, indicating an excellent linear correlation between the NMR response and the concentration of fats in the products.

Instrument repeatability was tested by measuring one sample ten times. Prior to each measurement the sample was placed in a heating block for 20 minutes.

Since the magnet is temperature stabilised at 40°C, precision is enhanced by conditioning the sample at 40°C prior to inserting it into the NMR magnet.

Conditioning samples prior to measurements aids repeatability, since the NMR signal is temperature sensitive.

In addition, some samples must be heated to melt the fat so it becomes visible to the NMR. The sample only needs to be returned to the block for a further 20 minutes if the measurement is going to be repeated on that sample.

Laboratory managers reported that the testing results exceeded their expectations, and the group purchased an MQC low-field, benchtop NMR analyser. Currently, about 90-95 percent of the laboratory's samples are run on the MQC instrument and the lab has increased its efficiency because of the rapid measurement capability.

Fig. 3 compares the number of analyses per hour and the skills required for the NMR technique versus that of Soxhlet.

According to Kulsum Jassat, sales manager for the UK and Europe: "Throughput has increased from 80 to 200 samples per day. With solvent extraction, each sample took approximately six hours from set-up to completion, or a rate of about one sample per hour.

"With NMR, it is usually less than one minute per sample after sample set-up."

She adds: "The company achieved a return on investment (ROI) within five months of purchasing the instrument."

The purchase of the NMR MQC instrument allowed the laboratory to replace four of its Soxhlet units. This significantly reduced the cost of purchase and disposal of solvents, increased available laboratory space, reduced running costs, achieved environmental benefits associated with reduced solvent usage, and facilitated better deployment of laboratory staff.

Since there is a slight calibration drift over about six months to a year, the laboratory will still need to recalibrate periodically against the primary technique or with set-up samples supplied by Oxford Instruments.

During installation, set-up samples can be made to physically capture and maintain the original calibration, saving time during instrument recalibration. They also save time by allowing calibrations to be transferred to other similarly configured Oxford MQC NMR instruments.

While not required by most potential users of NMR instruments, this particular contract lab chose to get its new fat measurement method accredited by the United Kingdom Accreditation Service (UKAS); the national accreditation body that uses a set of internationally agreed standards to assess organizations that provide certification, testing, inspection, and calibration services.

The lab developed a universal method using a sample blending process, sample conditioning to ensure the optimal temperature for reading by the temperature-sensitive NMR signal, followed by analysis using the MQC instrument. They used hundreds of data samples from Soxhlet and NMR-MQC to prove they were getting similar results within a degree of tolerance. This method has been UKAS-validated and fully accredited since 2009.[Page Break]

Fat content

NMR provides a fast and reliable method to increase throughput for quality control testing of fat content in foodstuffs. Although past history may have placed the venerable Soxhlet solvent extraction technique in the role of industry standard, for those seeking to eliminate quality control bottlenecks, the MQC NMR analyser is poised to take over the lead.