Alternatives to helium

In the face of rising prices for helium, Ed Connor reveals the more cost-effective options for gas chromatography.

Although there is now less concern about its availability, the price of helium is set to continue to increase, meaning that cost-effective alternatives are still attractive to analytical labs.

Many of the labs affected by the helium shortage of 2012-2013 have found that they are now able to source helium readily, although they have seen stark increases in price over the past two years.

One direct consequence of the helium shortage was the increase in information made available to chromatographers about alternative carrier gases and the resulting increase in the uptake of nitrogen and hydrogen as alternatives, owing to a change in perception relating to these gases.

Nitrogen is seen as a slow gas and is often overlooked as an alternative to helium, when in fact its use would be perfectly valid in a number of GC analyses. With a low optimal linear velocity of 8-14 cms-1 for nitrogen, compared with 25-33 cms-1 for helium (Fig. 1), analysis times will be increased if the analyst wants to maintain optimal performance. However, if there is enough resolution between peaks, it is possible to run samples at a higher average linear velocity. This means sacrificing some theoretical plates, which in practical terms means broadened peaks.

Hydrogen, on the other hand, is more efficient than helium at higher linear velocities, with an optimal linear velocity range of 38-45 cms-1, meaning that efficiency is improved compared with helium.

This improved efficiency at higher linear velocities potentially allows for increased sample throughput, however it is not always possible to analyse a sample at a higher linear velocity because of lack of peak resolution. Matching the linear velocity of hydrogen to that of helium should mean like-for-like analysis, with slightly improved carrier efficiency.

One concern that people have about working with hydrogen is its flammability, and placing hydrogen cylinders in the lab is a potential health and safety risk. Therefore the use of generators to produce hydrogen, and indeed nitrogen, is a cost-effective and safe method of gas production within the lab. Hydrogen is produced by the electrolysis of deionised water and is supplied to the GC on-demand. The hydrogen generator contains only a low volume of hydrogen at much lower pressure than cylinders, whilst being able to produce enough gas to supply a whole lab. Nitrogen is produced by the removal of oxygen, CO2 and hydrocarbons from compressed air by pressure swing adsorption using a carbon molecular sieve material.

A three-component alkane standard was run using a Shimadzu 2010 GC with a Restek RTX-1 column (30 m x 0.25 mm x 0.25 µm) using cylinder helium, and hydrogen and nitrogen produced by gas generators as the carrier gases.

The samples were run isothermally (170°C) using the three carrier gases at the same linear velocity (37.5 cms-1) to look at the effect on peak area, peak width, theoretical plate count and resolution.

Fig. 2 shows that the results obtained from this short run are very similar regardless of the carrier gas used.

Hydrogen and helium give almost identical results, with the peaks in the nitrogen run showing a little band broadening and although the linear velocity of all three carrier gases was set to 37.5 cms-1, the alkane peaks run with nitrogen carrier are slightly delayed compared with the peaks obtained using hydrogen and helium.

The data in Table 1 shows that all three carrier gases produced the same peak areas for each of the three compounds, so there was no effect of carrier gas on sensitivity.

As expected, however, running nitrogen at a high linear velocity has the effect of reducing efficiency, which is shown in the lower number of theoretical plates and therefore reduced resolution. The reduced efficiency (Table 1) has the effect of broadening the peaks, which is visible in the chromatogram and is shown in Fig. 2.

For analysis of samples such as the alkane mixture, which has plenty of resolution between peaks, there would be no reason why nitrogen and hydrogen cannot be used instead of helium.

Although use of nitrogen results in a reduction in theoretical plates, this does not impact on the overall results seen here. There is a strong case for using nitrogen for a number of GC analyses where high efficiency is not essential.

Hydrogen performed very similarly to helium, which demonstrates why many labs are now moving their analyses to hydrogen since they can get comparable chromatography, whilst using a cheap, readily available alternative to helium produced safely by generators. 

The price of helium is set to continue to rise so the results here show that nitrogen and hydrogen can be used as alternative carrier gases without loss of performance. Production of these gases by generators means that labs have a constant, on-demand supply of gas and are free from the annoyance of running out of gas in the night or having to move heavy cylinders around.

For more information visit www.scientistlive.com/eurolab

Ed Connor is GC product specialist at Peak Scientific